SETUP AND CALIBRATION, CONT.
SB-200 – (Type = 2)
If SB-200 is selected, the data will be transmitted in the following format:
<cr><s><xxxxxx><d><c><uu><m> ETX
Where:
cr =
Carriage Return
s =
Sign
xxxxxx.xxx = Weight
d =
Decimal point
c =
status
uu =
Units
m =
Mode
ETX =
End of TeXt
Rice Lake IQ355 – (Type = 3)
If Rice Lake IQ355 is selected, the data will be transmitted in the following format:
<stx><s>xxxxxxx<u><m><s><cr><lf>
AnDFV – (Type = 4)
If AnDFV is selected, the data will be transmitted in the following format:
<hdr1>,xxxxx<uu><cr><lf>
WI110 – (Type = 5)
If WI110 is selected, the data will be transmitted in the following format:
<m><sp><s>xxxxx<sp><uu><cr><lf>
Number Only – (Type = 6)
If number only is selected, the data will be transmitted in the following format:
xxxxxx<cr><lf>
Toledo Short – (Type = 7)
If Toledo Short is selected, the data will be transmitted in the following format:
<stx><swa><swb><swc>xxxxxx<cr><sum>
Where:
stx =
swa =, swb=, swc=
xxxxxx=
cr =
sum=
8200-M129-O1 Rev D 205 Installation & Technical
(hex 0D)
"-" = negative, " " (blank) = positive
(with leading zeros)
Embedded into weight (after weight dpp=0)
m = motion
o = overcap
e = weight not currently being displayed
tn, lb, l/o, oz, t, kg, g
G = Gross, N = Net
(hex 03) MUST terminate ALL serial commands
Start of TeXt (hex 02)
Status Bytes
Displayed Weight, Gross or Net Weight (Six Digits)
Carriage Return (hex 0D)
Checksum Character