SETUP AND CALIBRATION, CONT.
Dual-Point without Zero (False Zero) Calibration
IMPORTANT! The functions of numeric keys are replaced by using the UNITS/LEFT
ARROW and the /UP ARROW keys. The cursor location is identified by the
blinking character and can be advanced to the left to the next position by pressing
the UNITS/LEFT ARROW key. Pressing the /UP ARROW key will change the
blinking character to the next value.
CAL1= – First Calibration Weight
The display will show CAL1=0. This is the first of two calibration steps. This weight is
the TEST WEIGHTS / TEST LOAD.
• Place the weights on the scale platform.
• Using the UNITS/LEFT ARROW and the /UP ARROW keys, input the value of the
calibrated test weights / test load, then press the TARE/ENTER key. NOTE! When
entering values for CAL1=, the digits start displaying on the right side of the display
and proceed to the left. When large values are used (more than 3 digits), the CAL1=
prompt will automatically scroll off the left side of the display to show the additional
digits on the right as they are entered.
• Starting at the left and advancing right, a series of dashes will appear on the display.
The dashes will stay on the display momentarily, then disappear, after which the
display will show: CAL2=.
CAL2= - Second Calibration Weight
The display will show CAL2=0. This is the second of two calibration steps.
• Remove the weights on the scale platform, then press the NET/GROSS key.
• Starting at the left and advancing right, a series of dashes will appear on the display.
The dashes will stay on the display momentarily, then disappear, after which the display
will show: Sio?.
Single-Point for Span Only (Last Zero) Calibration
CAL1= – First Calibration Weight
The display will show CAL1=0. This is the first of two calibration steps. This weight is
the TEST WEIGHTS / TEST LOAD.
• Zero the scale, then place the weights on the scale platform.
• Using the UNITS/LEFT ARROW and the /UP ARROW keys, input the value of the
calibrated test weights / test load, then press the TARE/ENTER key. NOTE! When
entering values for CAL1=, the digits start displaying on the right side of the display
and proceed to the left. When large values are used (more than 3 digits), the CAL1=
prompt will automatically scroll off the left side of the display to show the additional
digits on the right as they are entered.
• Starting at the left and advancing right, a series of dashes will appear on the display.
The dashes will stay on the display momentarily, then disappear, after which the
display will show: CAL2=.
CAL2= - Second Calibration Weight
The display will show CAL2=0. This is the second of two calibration steps.
• Remove the weights on the scale platform, and then press the ZERO key.
• The display will advance to Sio?.