SETUP AND CALIBRATION, CONT.
b= (Break Range)
Press the TARE/ENTER key to show the current setting for the break range. The break range
is a number from 1 to 255 that corresponds to the number of division change to break out of
the filtering.
To accept the value displayed, press the TARE/ENTER key, otherwise, use the /UP ARROW
and UNITS/LEFT ARROW keys to enter a new setting and then press the TARE/ENTER key to
save it. Note that selecting 0 disables this feature.
h= (High Frequency Filter Value)
Press the TARE/ENTER key to show the current setting for the high frequency filter level. The
filter level is a number from 0 to 99 that corresponds to the level of filtering with 99 being the
greatest filtering and 1 the least.
To accept the value displayed, press the TARE/ENTER key, otherwise, use the /UP ARROW
and UNITS/LEFT ARROW keys to enter a new setting and then press the TARE/ENTER key to
save it. Note that selecting 0 disables this feature.
P= (Preliminary Filter)
Press the TARE/ENTER key to show the current setting for the high frequency filter level. The
preliminary filter is a number from 0 to 99 that corresponds to the number of weight samples
that will be filtered without breaking (regardless of what b= is set to).
To accept the value displayed, press the TARE/ENTER key, otherwise, use the /UP
ARROW and UNITS/LEFT ARROW keys to enter a new setting and then press the
TARE/ENTER key to save it. Note that entry of a 0 or 1 disables this feature.
Sr= (Sample Rate)
Press the TARE/ENTER key to show the current value. If the setting displayed is acceptable,
press the TARE/ENTER key again to save it. Otherwise, use the /UP ARROW and
UNITS/LEFT ARROW keys to enter a new setting and then press the TARE/ENTER key to
save it. Allowable values are: 1 through 100.
UnS= (Motion Range)
Press the TARE/ENTER key to view the current setting for the range of motion detection. If
the displayed value is acceptable, press the TARE/ENTER key to save it. Otherwise, use the
/UP ARROW and UNITS/LEFT ARROW keys to enter the new range (the number of
divisions of change permitted before indicating unstable), then press the TARE/ENTER key to
save the new setting. Allowable range values are: 0 through 99 divisions.
SC= (Stable Count)
Press the TARE/ENTER key to view the current setting for the number of consecutive stable
weight readings before indicating stable weight. This helps filter weight readings for stability
for use with Auto Print on Balance, or and anything trying to capture stable weight.
If the displayed value is acceptable, press the TARE/ENTER key to save it. Otherwise, use
the /UP ARROW and UNITS/LEFT ARROW keys to enter a new setting and then press the
TARE/ENTER key to save it. Allowable values for the stable count are: 3 through 255.