Rotors For The Accuspin™ 1 - Fisher Scientific accuSpin 1 Operating Instructions Manual

Table of Contents

Advertisement

Rotors and accessories
Rotors for the accuSpin™ 1
Table 1:
Rotor designation
With bucket
Order no.
Maximum permissible load [ g ]
Maximum speed n
[ rpm ]
max
Maximum RCF value at n
Radius (max.)
[ cm ]
Acceleration / deceleration time [ s ]
Aerosol-tight
Autoclavable
* Checked by CAMR, Porton-Down, UK
14
Round bucket
400 ml
7500 3451
4 x 600
4000
3345
max
19.0
60 / 55
yes*
(with cap 75007584)
no
Swinging Bucket Rotor 7500 3450
Rectangular bucket
150 ml
7500 3452
4 x 510
4000
3220
18.1
60 / 55
yes
no
Carrier for
microplates
7500 3453
4 x 450
4000
2773
15.5
60 / 55
no
no