Worn And New Stator Pads - Beckman Coulter Airfuge Instructions For Use Manual

Air driven ultracentrifuge
Hide thumbs Also See for Airfuge:
Table of Contents

Advertisement

Figure 3.1 Worn and New Stator Pads
1
1. Worn Stator Pads
5
Place the rotor on the stator pad and shut the instrument door.
6
Set the
• The dial indicates up to 5 hours of run time in 5-minute increments.
— (In some models, the dial indicates 5 minutes of run time in 5 second increments.)
a. For longer runs, set the timer to
b. For runs shorter than 30 minutes, turn the dial past the 30 minute point, then back to the
required time.
• ACR-90 rotor—set the run time for 10 minutes to prevent significant evaporative loss.
7
Secure the instrument door by turning the pressure regulator knob to the right (clockwise),
pushing down, until the air pressure indicated on the
required speed. Optimal run speed is achieved at 30 psig (207 kPa). (This is the only
recommended operating air pressure for the ACR-90 rotor.) Note that some low-speed rotor
vibration is normal at the beginning of the run.
AF-IM-13AC
2
2. New Stator Pads
dial for the required run time.
TIME
.
HOLD
PRESSURE
Operation
Starting the Run
gauge brings the rotor to the
3
3-3