Decontamination; Location Of Time-Delay Knob - Beckman Coulter Airfuge Instructions For Use Manual

Air driven ultracentrifuge
Hide thumbs Also See for Airfuge:
Table of Contents

Advertisement

WARNING
Disconnect the power cord and the air supply.
Figure 5.7 Location of Time-Delay Knob
1. Time-Delay Knob
1
Close the Airfuge door and turn the instrument on its side.
2
Turn the time-delay knob counterclockwise to shorten the delay time, or clockwise to lengthen
the delay time.
3
Reconnect the power cord and the air supply and perform a trial run to check the new setting.

Decontamination

If the instrument and/or accessories are contaminated with radioactive or pathogenic solutions,
follow appropriate decontamination procedures as determined by your laboratory safety officer.
Refer to Chemical Resistances (publication IN-175) or contact Beckman Coulter Field Service to
ensure that the decontamination method does not damage any part of the instrument (or
accessories).
AF-IM-13AC
1
5
Maintenance
Decontamination
5-11