Configure Network Connections - Viavi mA-1302 Operation Manual

Axie
Table of Contents

Advertisement

Configure Network Connections

The Chassis can be connected to a network using the Ethernet Connector for a wired LAN
connection. Chassis network connections are managed from the Chassis Networking
Page (see
IMPORTANT NETWORK CONFIGURATION INFORMATION
Read this section in its entirety before making any changes to the CMP System's
network settings. Failure to comply with all information provided in this section may
cause system network communication problems. If you are not familiar with the
terminology in this section, VIAVI recommends that you contact your network
administrator before proceeding.
The information in this section only applies to VIAVI CMP AXIe Modules; refer to OEM
user documentation for non-VIAVI products.
All VIAVI modules in a CMP System must be configured to use the same network
mode of operation (DHCP or Static).
The IP address of each module in a CMP System must be within range of the other
modules in the system or the modules will not be able to communicate with each other.
In the event you lose IP system connectivity, perform the Chassis IP reset procedure
("How to Reset the Chassis IP Address" on page
communication. Once you restore connectivity with the Chassis, refer to the Chassis
Networking page for the IP address of the modules located in the Chassis. The
Chassis Networking Page is also used to reset the IP addresses for all CMP modules
located in the Chassis.
DEFAULT STATIC IP NOTICE
The Chassis ships from the factory with an assigned Static IP Address. Do not
connect the Chassis to your local area network (LAN) without contacting your IT
department to determine if the Chassis' default IP Address is already in use. If the
Chassis' default IP Address is already in use by a device on the network, connecting
the Chassis to the network will cause a network conflict.
It is recommended that the Chassis be reconfigured to use a Static IP address
assigned by your IT department before connecting the Chassis to a LAN.
Rev. D0
May 2020
"Chassis Networking Page" on page
mA-1302/mA-1305 AXIe Chassis Operation Manual
Chapter 5 mA-1302/mA-1305 Control and Operation

Configure Network Connections

4-17).
5-15) to restore system
Page 5-9