NM2C
4.20.11 Slide the display onto the mounting rod, oriented so that the
4.20.12 Tighten the display mounting clamp screws (accessible through
4.20.13 Install the remote display and Datagrip cables in the cable clamps
4.20.14 Install a tie strap to secure the remote display and Datagrip cables
4.20.15 Connect the remote display and Datagrip cables to their ports on the
4.20.16 Restore power to the machine and verify operation of the remote
4.20.17 Perform the PMS Procedure given in Section 6.
Rev. J
RETURN TO THIS MANUAL'S TABLE OF CONTENTS
REPLACEMENT PROCEDURES (continued)
mounting clamps in the display assembly slide past the stop pins on
the rod - until the display is in the correct position on the rod.
clearance holes in the back cover of the display) until the display has
the desired amount of friction on the mounting rod.
that were previously removed, and reinstall their screws along the
display arm, bellows box and ventilator box. Ensure that the display
arm has full range of motion with no binding caused by the cables.
together.
underside of the ventilator box.
display and Datagrip by exercising the display selections.