REPLACEMENT PROCEDURES (continued)
4.18.9
Slide the processor assembly into the ventilator chassis and secure it
with the two screws that were previously removed.
4.18.10 Connect the O.R. DATA MANAGER ribbon cable to J5 on the
processor board.
4.18.11 Reinstall the back panel of the ventilator box and secure it with the
six screws that were previously removed.
4.18.12 Reconnect the Datagrip and remote display cables, and restore any
serial interface or printer connections.
4.18.13 Restore power to the machine and observe the Power-Up Diagnostic
display (see Section 2) to verify that the replacement processor is
working properly.
4.18.14 Access the secondary service screen and enter the machine serial
number and vaporizer configuration into memory.
Draeger Medical, Inc. Technical Service Department for access
instructions if necessary.
4.18.15 Perform the PMS Procedure given in Section 6.