ADJUSTMENT AND CALIBRATION PROCEDURES (continued)
SV20316
RECEPTACLE
ON INTERFACE
PANEL
SENSOR
CORD
CONNECTOR
OXYGEN MONITOR SERVICE SCREEN
CURRENT CELL A: 366
CURRENT CELL B: 371
ZERO CALIBRATION PROCEDURE:
- REMOVE O2 CELL FROM SENSOR HOUSING
- LET CURRENT CELL VALUES STABILIZE
- SELECT "STORE ZERO" KEY
PRESS TRIGGER TO ENTER CALIBRATION VALUES
- REINSTALL OXYGEN CELL IN SENSOR HOUSING
STORED ZERO CELL A: 240
STORED ZERO CELL B: 241
Figure 5-5: OXYGEN MONITOR SERVICE SCREEN