Dräger Narkomed 2C Technical & Service Manual page 145

Anesthesia system
Table of Contents

Advertisement

NM2C
PM Certification Procedure for Narkomed 2B/2C/GS Anesthesia System
7.
Proceed with PM Certification procedure in accordance with SP00152, SP00150 or
SP00169. If any tests fail refer to the "Failure Codes" listing to determine correct
certification level starting point. Failure codes shall be documented on the
"RECOMMENDATIONS / GENERAL COMMENTS" section of the PM Certification
form and on the Executive Summary. If a test fails that has not been identified by the
"Failure Codes" list, consult with Draeger Medical, Inc. to assess the proper certification
level.
8.
Based on the "EQUIPMENT CONDITION" inspect the machine for any "NAD
RECOMMENDATIONS" that would apply. Use the Narkomed 2B, 2C or GS section of
the "RECOMMENDATION GUIDELINES INDEX". Note all applicable NAD
recommendations on the Executive Summary. NOTE: If using a carbon form, indicate
the
Equipment
"RECOMMENDATIONS / GENERAL COMMENTS" section of the form.
9.
Determine the correct certification level of the machine based on the combined lowest
common denominator of "Equipment Conditions" and "Failure Codes". If the machine
is at least conditionally certified fill out the "PM CERTIFICATION" label. Check the
box(s) on the validation label where appropriate. Write the month and year, (three
months from date of PM Certification) next to "NEXT VISIT DUE:" If certification level
is "D", machine shall not receive a "PM CERTIFICATION" label. Any machine not
receiving a PM Certification label shall receive a "WARNING NOT CERTIFIED" label.
This label shall be placed at a prominent location on the left side of the machine after
all other previous PMS and Vigilance Audit Validation labels have been removed.
10. In the "CERTIFICATION LEVEL" section of the PM Certification form, record the last
visit certification level, the current certification level and the next visit due month and
year, (three months from date of PM Certification) in the spaces provided.
11. If applicable, remove the previous PM CERTIFICATION VALIDATION label and attach
the new label in a prominent location on the rear of the anesthesia machine.
12. Check the appropriate boxes on the "PM CERTIFICATION NOTICE" label, (part #
S0100011). If the machine is not certified, the last box of this notice label shall be
marked. Attach this notice to the flow shield of the anesthesia machine.
13. Have the customer sign each PM Certification form or the Executive Summary, and
review the equipment conditions and the recommendations with the customer.
14. Return top copy to Draeger Medical, Inc. Service Department, keep middle copy for
service organization records, give bottom copy to customer.
Rev. T
RETURN TO THIS MANUAL'S TABLE OF CONTENTS