Dräger Narkomed 2C Technical & Service Manual page 49

Anesthesia system
Table of Contents

Advertisement

NM2C
4.5.19
Place the plexiglass cover over the gauges and flow tubes, and ensure
that the cover is fitted properly over the flow control valves.
4.5.20
Place the knob guard over the flow control valves and install its two
retaining screws.
4.5.21
Install the oxygen flow control knob and tighten its setscrews. If the
knobs are installed properly, their labels will be straight when the
knobs are against their clockwise stops.
4.5.22
Replace the front plate at the top of the plexiglass cover and secure
it with the hardware that was previously removed.
4.5.23
Replace the table top and its retaining screws.
4.5.24
Connect the pipeline hoses.
4.5.25
Perform the PMS Procedure given in Section 6.
RETURN TO THIS MANUAL'S TABLE OF CONTENTS
RETURN TO CD-ROM TABLE OF CONTENTS
REPLACEMENT PROCEDURES (continued)
4-19