Dräger Narkomed 2C Technical & Service Manual page 21

Anesthesia system
Table of Contents

Advertisement

NM2C
START
IS
BREATHING PRESSURE
MONITOR WORKING
PROPERLY?
CHECK PILOT
LINE CONNECTION
AT BOTH ENDS
N
IS
PILOT LINE
CONNECTED?
CHECK INTERNAL
TUBING FROM
INTERFACE PANEL
TO TRANSDUCER
ON PROCESSOR ASM.
IS
BREATHING PRESSURE
MONITOR WORKING
PROPERLY?
REPLACE
PROCESSOR ASM.
AS OUTLINED IN
SECTION 4
IS
BREATHING PRESSURE
MONITOR WORKING
PROPERLY?
RETURN TO THIS MANUAL'S TABLE OF CONTENTS
TROUBLESHOOTING GUIDE (continued)
GUIDE 2: Loss of Breathing Pressure Monitor
Y
N
Y
Y
N
PERFORM A
COMPLETE PMS
ON UNIT
Y
N
REINSTALL ORIGINAL
PROCESSOR ASSEMBLY
3-5
RETURN TO CD-ROM TABLE OF CONTENTS
DOES UNIT
PASS PMS?
N
CONTACT NAD
SERVICE DEPT.
Y
UNIT IS
FUNCTIONAL