PMS PROCEDURE (continued)
6.30 OXYGEN FLUSH VALVE
6.30.1
Press and release the O
6.30.2
Does the flow of oxygen stop immediately? __ (Y)
6.30.3
Close the APL valve.
6.30.4
Connect a 12-inch hose to the inspiratory valve.
6.30.5
Set the Man/Auto selector to BAG.
6.30.6
Occlude the bag mount.
6.30.7
Insert the sensor from a calibrated O
on the inspiratory valve.
6.30.8
Close all flow control valves.
6.30.9
Press the O
( )
6.30.10
What is the O
6.30.11
Remove the O
6.30.12
Remove the test minute volumeter (P/N 2212300) from the absorber and
connect it to the common gas outlet, using Fresh Gas Outlet Volume
Test Device (P/N S010158).
6.30.13
Press and hold the O
by 4.
( )
6.30.14
What is the oxygen flush flow rate? ___l/min. (45-65)
6.30.15
Remove the test minute volumeter and test fixture, and reconnect the
fresh gas hose.
6.30.16
Turn the System Power switch to ON.
RETURN TO THIS MANUAL'S TABLE OF CONTENTS
RETURN TO CD-ROM TABLE OF CONTENTS
FLUSH button.
2
FLUSH button.
2
concentration after the value stabilizes?___%O
2
Med sensor and install the plug.
2
FLUSH button for 15 seconds; multiply the value
2
6-56
Med into the valve dome adapter
2
NM2C
(97-100)
2
Rev. L