Dräger Narkomed 2C Technical & Service Manual page 173

Anesthesia system
Table of Contents

Advertisement

NM2C
6.12.16
Open the corresponding cylinder valve and set the System Power switch
to ON.
6.12.17
Set the corresponding flow to 4 l/min.
( )
6.12.18
On the second test gauge, what is the regulator output pressure? ___
PSI (40 - 49 for N2O), (43 - 49 for Air)
6.12.19
Depressurize the gas circuit.
6.12.20
Remove the test gauge and replace the plug in the TEE fitting.
( )
6.12.21
Repeat Steps 6.12.13 thru 6.12.20 for all other gases except CO
Rev. U
RETURN TO THIS MANUAL'S TABLE OF CONTENTS
RETURN TO CD-ROM TABLE OF CONTENTS
PMS PROCEDURE (continued)
6-27
circuits.