ADJUSTMENT AND CALIBRATION PROCEDURES (continued)
5.3.17
The nitrous oxide flow should stop when the flow of oxygen is
between 250 and 400 ml per minute.
5.3.18
Close the N
to STANDBY.
5.3.19
Reconnect the pipeline hoses.
5.3.20
Replace the flowmeter housing rear cover.
5.3.21
Perform the PMS Procedure given in Section 6.
RETURN TO THIS MANUAL'S TABLE OF CONTENTS
RETURN TO CD-ROM TABLE OF CONTENTS
O flow control valve and turn the System Power switch
2
5-10
NM2C
Rev. C