Warnings/Error Signals - Otto Bock Kenevo 3C60 Instructions For Use Manual

Hide thumbs Also See for Kenevo 3C60:
Table of Contents

Advertisement

Appendices

13.2.2 Warnings/error signals

Error during use
Beep signal
10 x long
30 x long
Error while charging the product
LED on
LED on
Battery charger
power
battery
connected to
supply
charger
product
No
32 | Ottobock
Vibration signal
1x long at interval of
approx. 5 seconds
3 x long
5 x long
10 x long
1x long, 1x short repeated
every 3 seconds
Error
Country-specific plug adapter not
fully engaged on power supply
Non-functional socket
Defective power supply
Event
Overheated hydraulic unit Reduce activity.
Battery
charge
level
under 35%
Battery
charge
level
under 20%
Battery
charge
level
under 15%
After the beep and vibra­
tion signal, the product
switches to empty battery
mode and then switches
off.
Severe error
For example, a sensor is
not
operational,
valve
drive failure or AXON
tube adapter not connec­
ted
Possibly no switching into
safety
mode
(see
Page 28).
Resolution
Check whether the country-specific
plug adapter is fully engaged on
the power supply.
Check socket with another electric
appliance.
The battery charger and power
supply must be inspected by an
authorised
Centre.
Required action
Charge battery soon.
Charge battery immediately,
since the product will be
switched off after the next
warning signal.
Charge the battery.
Restricted walking possible.
The possible change in flex­
ion/extension resistance must
be observed.
Attempt to reset this error by
connecting/disconnecting the
battery charger. The battery
charger must remain connec­
ted for at least 5 seconds
before it is disconnected.
If the error persists, it is no
longer permissible to use the
product. The product must be
inspected by an authorised
Ottobock Service Centre.