CAUTION
Overloading due to unusual activities
>
Falling due to unexpected product behaviour as the result of malfunction.
>
Falling due to breakage of load-bearing components.
>
Skin irritation due to defects on the hydraulic unit with leakage of liquid.
► The product was developed for everyday use and must not be used for unusual activities. These unusual activ­
ities include, for example, extreme sports (free climbing, paragliding, etc.).
► Careful handling of the product and its components not only increases their service life but, above all, ensures
the patient's personal safety!
► If the product and its components have been subjected to extreme loads (e.g. due to a fall, etc.), then the
product must be inspected for damage immediately. If necessary, forward the product to an authorised
Ottobock Service Centre.
CAUTION
Mechanical stress during transport
>
Falling due to unexpected product behaviour as a result of a malfunction.
>
Falling due to breakage of load-bearing components.
>
Skin irritation due to defects on the hydraulic unit with leakage of liquid.
► Only use the transport packaging for transportation.
CAUTION
Signs of wear to system components
Falling due to damage or malfunction of the product.
► In the interest of the patient's safety and in order to maintain operating reliability, the product should be ser­
viced at regular intervals.
NOTICE
Improper product care
Damage to the product due to the use of incorrect cleaning agents.
► Only clean the product with a damp cloth and mild soap (e.g. 453H10=1 Ottobock DermaClean).
INFORMATION
When using exoprosthetic knee joints, servomotor, hydraulic, pneumatic or brake load dependent control func­
tions can cause movement noise. This kind of noise is normal and unavoidable. It generally does not cause any
problems. If movement noise increases noticeably during the lifecycle of the knee joint, the knee joint should be
inspected by an authorised Ottobock Service Centre immediately.

4.4 Information on the Power Supply/Battery Charging

CAUTION
Charging the prosthesis without taking it off
Falling due to unexpected behaviour of the prosthesis because of changed damping behaviour.
► Instruct the patient that the prosthesis must be taken off before it is charged.
CAUTION
Charging the product with damaged power supply unit/charger/charger cable
Falling due to unexpected behaviour of the product caused by insufficient charging.
► Check the power supply unit, charger and charger cable for damage before use.
► Replace any damaged power supply unit, charger or charger cable.