7.5.1 Attaching the cosmetic cover
CAUTION
Use of talcum
Risk of injury, damage to the product due to lack of lubrication
► Do not use talcum on the product or other prosthesis compon
ents.
Recommended combinations
Foam cover
► After affixing the cosmetic elements, verify proper functioning of
the prosthesis.
► To reduce friction and to eliminate noise, apply silicone spray dir
ectly onto the contact surfaces of the cosmetic foam cover.
8 Use
8.1 Information for use
NOTICE
Mechanical overload
Impaired functionality due to mechanical damage
► Check the product for damage prior to each use.
► Do not use the product if its functionality has been impaired.
► Take any necessary measures (e.g. repair, replacement, inspec
tion by the manufacturer's customer service, etc.).
8.2 Cleaning
CAUTION
Use of unsuitable cleaning agents or disinfectants
Impairment of functionality and damage due to incorrect cleaning
agents or disinfectants
► Clean the product only according to the instructions given in this
section.
3R6, 3S106
► For the prosthesis, observe the cleaning instructions for all pros
thetic components.
INFORMATION
► Clean the product if it gets dirty.
► For disinfection, only use disinfectants that do not affect the
materials of the product. For further information, please consult
the manufacturer.
► Observe the cleaning instructions for all prosthetic components.
1) Clean the product with a damp, soft cloth.
2) Dry the product with a soft cloth.
3) Allow to air dry in order to remove residual moisture.
9 Maintenance
CAUTION
Failure to follow the maintenance instructions
Risk of injuries due to changes in or loss of functionality and dam
age to the product
► Observe the following maintenance instructions.
► Arrange regular maintenance intervals with the patient depending
on the level of use.
► Following an individual period for the patient to get accustomed
to the prosthesis, check the settings of the prosthetic joint and
adapt them to the patient's requirements again as needed.
► The prosthetic components should be inspected after the first 30
days of use.
► Inspect the entire prosthesis for wear during normal consulta
tions.
► Conduct annual safety inspections.
21