REVEXmax - User's Manual
4.6.2

Three-phase and hard-wired appliances

Optional accessories are intended for leakage current measurement of both three-
phase and hard-wired appliances. Three-phase adapter must be used for measure-
ment of three-phase appliances with movable power supply cord. Current clamp
must be used for measurement of hard-wired appliances.
It is necessary to plug in tested appliance by means of three-phase adapter and to
select required measuring method by means of three-phase adapter's function switch
before measurement (see chapter 7.4.4).
Procedure:
Press the [4 I
method. The symbol
Connect either optional current clamp or three-phase adapter to the termi-
nal 8.
If you use current clamp then enclose with its jaws either protective conductor
or all current carrying (L+N) conductors of tested appliance.
Switch on the tested appliance's mains switch.
Press the [TEST] key.
The last measured value is displayed after finishing the measurement.
Switch off the tested appliance's mains switch.
c 2019 ILLKO, s.r.o.