REVEXmax - User's Manual
7 Additional information
Figure 8: Warning message: Current >10.00 mA
Measuring - time limit
If the time of measuring exceeds the allowed limit (according to the appliance's
current - see chapter 2.1), power supply is disconnected from the Test Socket 1, then
an acoustic signal sounds and the display shows the warning message (Figure 9).
Figure 9: Warning message: Measuring - time limit
Cannot be measured (calibration error / damaged data)
The device automatically checks the adjustment constants and other data. If the
data is damaged then the measurement cannot be performed. An acoustic signal
sounds and the display shows the warning message (Figure 10 or Figure 11). In
both cases, the device must be sent for repair to the manufacturer.
Figure 10: Warning message: Cannot be measured (calibration error)
Figure 11: Warning message: Cannot be measured (damaged data)