REVEXmax - User's Manual
7.7

Recurrent tests and testing after repair

The REVEXmax is powered with a mains voltage of 230 V / 50 Hz and therefore
all recurrent tests and testing after repair should be done in accordance with local
regulations.
Insulation resistance between LN and PE in the REVEXmax must not be measured
because of varistors in EMC filter.
The protective conductor must produce a test current approx. 1.5 mA. If leakage
current is measured (instead of insulation resistance measurement), the value must
remain below:
3.5 mA if the Direct or Differential leakage current measurement has been
chosen
7 mA if the Equivalent leakage current has been chosen
c 2019 ILLKO, s.r.o.
www.illko.cz
7 Additional information
Page 61