REVEXmax - User's Manual
4.8
Apparent power, mains voltage and current consump-
tion - Power
4.8.1

Power, mains voltage and current consumption in Test Socket

If test voltage is present in Test Socket 1 – it is indicated by the (TEST)
indicator, then the tested equipment starts to run after its mains switch
is switched on! Take into consideration all safety warnings referred to
by the DUT!
Do not touch the exposed conductive parts of the tested object after
the mains switch is switched on – RISK OF ELECTRIC SHOCK!
WARNING
Test Socket 1 of the REVEXmax must not be used for continuous
power supply of appliance!
Procedure:
Press the [9 Power] key and release it.
Connect the tested appliance: plug in its mains plug into Test Socket 1.
Press the [TEST] key.
Switch on the tested appliance's mains switch.
The HOLD function can be used (press the [ ] key).
Switch off the tested appliance's mains switch.
Finish the measurement (see chapter 4.2.1).
If the measured value of power is higher than the highest measurable value the
symbol >>>> appears.
c 2019 ILLKO, s.r.o.
www.illko.cz
4 Measurements