User's Manual
4.2.2. Protective class I three-phase and hard-wired machines
Step 1
Press the ∆U/R
•
and release it again to select 10A earth bond function).
Either result of last measurement or „---" will be displayed. The set value of cross-
•
section of protective conductor will be displayed, too.
∆
Step 2
If you need to change timer setting, proceed as follows:
Press the MENU key; the main menu is opened.
•
Use ▲ and ▼ keys to select submenu „Set-up" and press ENTER.
•
Use ▲ and ▼ keys to select item „Timer RPE 10A" and press ENTER.
•
Use the DEL key to erase current setting.
•
Copyright © 2004 ILLKO, s.r.o.
SONDE
R
R
PE
ISO
R
0
PE
200mA
∆
U/R
1
PE
10A
2
R
ISO
3
I
LEAK
4
I
PE
10A key and release it to select 10A voltage drop function (press
PE
U10A
1,0mm2
---
PRÜFDOSE / TEST SOCKET
TEMP.
I
/
LEAK
5
I
PE INV.
ESC 2051
I
6
DIFF.
DEL
I
7
DIFF. INV.
LOAD
DEL
P
8
VA
START
SAVE
CAL
9
MENU
Fig. 4
16
ESC
ENTER
ESC 2051