4.2.2. Protective class I three-phase and hard-wired machines

Step 1
Press the ∆U/R
and release it again to select 10A earth bond function).
Either result of last measurement or „---" will be displayed. The set value of cross-
section of protective conductor will be displayed, too.
Step 2
If you need to change timer setting, proceed as follows:
Press the MENU key; the main menu is opened.
Use ▲ and ▼ keys to select submenu „Set-up" and press ENTER.
Use ▲ and ▼ keys to select item „Timer RPE 10A" and press ENTER.
Use the DEL key to erase current setting.