RSLA/RELA scale installation drawing

Dimensions and tolerances in mm
Alignment dowels, maximum spacing 350
Increasing count direction
NOTE: Head can be mounted in either orientation.
Encoder count direction is dependent only on scale orientation.
5
NOTE: Adhesive mounted scale should not be reused after installation.
*
 RELA shown. RSLA installation is the same but the scale has no chamfers.
 
When scale is to be mounted vertically, position the dowels so that the datum edge is supported.
 Double-sided adhesive tape is included with all scale lengths.
RESOLUTE FS with BiSS
Safety encoder system installation guide and safety manual
®
*  (Adhesive mounted RELA shown)
Measuring length, ML = (L-10)
Optical centreline
L
0.5
0.05/50
F = axis of motion
Adhesive datum clamp
12
RGG-2, 2-part epoxy, usually positioned
10
coincident with user selected scale datum
Original instructions
Datum edge
F
Double-sided adhesive tape
5
19