Manual Calibration - Graco SDP8 Series Installation And Operation Manual

Preset metered dispense valve
Table of Contents

Advertisement

Pulse FC Mode
5.
Dispense exactly one quart or one liter
of fluid into a one quart or one liter clean,
calibrated, volumetric flask.
NOTE: During calibration dispense the
metered dispense valve will not display
the volume dispensed. The volume
dispensed is only determined by the
flask measurement. The following
screen displays during the calibration
dispense.
F
. 60
IG
6.
When exactly one quart or one liter of
fluid is dispensed into the flask use the
center ENTER button on the keypad to
select END. A screen displaying the
quantity of fluid dispensed displays (F
61 shows an example of the dispensed
volume screen).
F
. 61
IG
7.
Press the center ENTER button on the
keypad to save the new calibration
factor.
30
8.
After selecting END the Main Menu
screen displays.
F
. 62
IG

Manual Calibration

NOTE: This Manual Calibration procedure is
used when the k-factor is known and a simple
adjustment of the displayed k-factor is
needed.
1.
Follow steps 1- 3 of the Calibration
instructions (page 29).
2.
Use the RIGHT ARROW button on the
keypad to highlight +/- and press the
center ENTER button on the keypad to
select the +/- option (F
.
IG
F
. 63
IG
. 63).
IG
3A6673B