8 Adjustment

As the acceleration value due to gravity is not the same at every location on earth,
each balance must be coordinated - in compliance with the underlying physical
weighing principle - to the existing acceleration due to gravity at its place of location
(only if the balance has not already been adjusted to the location in the factory). This
adjustment process must be carried out for the first commissioning, after each
change of location as well as in case of fluctuating environment temperature. To
receive accurate measuring values it is also recommended to adjust the balance
periodically in weighing operation.
 Observe stable environmental conditions. A warming up time (see chapter 1)
is required for stabilization.
 Ensure that there are no objects on the weighing plate.

8.1 Models with external weight (KERN ALS/PLS)

The adjustment is locked for verified balances.
Weight value of the required adjustment weight see chpt. 1 "Technical
specifications"

8.1.1 Adjustment with recommended adjustment weight (factory setting)

 Ensure that there are no objects on the weighing plate. Press the CAL key
 Wait until the weighed value for the required adjustment weight appears
flashing.
 During the flashing display put the required adjustment weight carefully in the
center of the weighing plate.
The flashing display disappears.
After successful adjustment the balance automatically returns to weighing mode.
 Take away adjustment weight
ALJ/ALS/PLJ/PLS-BA-e-1943