2.

BEFORE USING THE ANALYSER EVERY TIME:

Check the water trap is empty and the particle filter is not dirty:
-
To empty water trap, unscrew the red screw plug and re-tighten once it is
empty.
-
To change the particle filter, remove protective rubber cover, slide the water
trap unit from the analyser, remove the particle filter from its spigot and replace.
Reconnect the water trap unit and rubber protective cover.
Connect the flue probe hose to the analyser's flue gas inlet and connect the flue
probe's temperature plug to the T1 socket – check the plug's orientation is correct -
see Page 6.

2.1 FRESH AIR PURGE

Position the flue probe in fresh air, then press
starts and the analyser auto-calibrates. When complete:
Select "Ratio" on the dial.
Select "O
/Eff" on the dial.
2
FRESH AIR
PURGE
REMOVE
PROBE
FROM FLUE
PRESS
This message indicates that the analyser needs to be reset in fresh air. To do so,
ensure that the analyser is in fresh air and press
To perform a manual 'Gas Zero', select 'Ratio' on the dial, hold down the
key and you will see the message above.
In fresh air the CO reading should be zero.
In fresh air the O
2
/
. The analyser's pump
reading should be 20.9% ± 0.3%.
/
.
Page 8