Where To Send Your Analyser; Cold Weather Precautions - Kane 456 Manual

Flue gas analyser with direct co2 measurement
Table of Contents

Advertisement

12.3 WHERE TO SEND YOUR ANALYSER

Northern Customer Service
Kane International Ltd
Gibfield Park Avenue
Atherton,
Manchester
M46 0SY, UK
t: +44 (0) 1942 873434
f: +44 (0) 1942 873558
e: nservice@kane.co.uk

13. COLD WEATHER PRECAUTIONS

It is important you keep your flue gas analyser in a warm place overnight
Electronic devices that become really cold, by being left in a vehicle
overnight, suffer when taken into a warm room the next morning.
Condensation may form which can affect the analyser's performance &
cause permanent damage
Electrochemical sensors used in flue gas anlysers can be affected by
condensation or water being sucked into the analyser, as the small
apertures on top of sensors can become blocked with water, stopping
sensors seeing flue gas. When this happens, oxygen or carbon dioxide
reading will display as "—" & sensors may be permanently damaged
If you think that your analyser is affected by condensation or water
ingress, it may be possible to rectify the problem yourself. Simple leave
the analyser running in a warm place, with the pump 'ON' sampling fresh
air for a few hours (use mains adapter/battery charger if needed). If, after
doing this, you still experience problems please contact our Service
Centres.
Southern & International Customer Service
Kane International Ltd
Kane House, 11 Bessemer Road
Welwyn Garden City
Hertfordshire
AL7 1GF, UK
t: +44 (0) 1707 384834
f: +44 (0) 1707 384833
e: sservice@kane.co.uk
Page 43

Hide quick links:

Advertisement

Table of Contents
loading

Table of Contents