14. ANALYSER SPECIFICATION

(NOTE: MAY BE SUBJECT TO CHANGE)
Parameter
Temp Measurement
Flue Temperature
Inlet Temperature
(Internal sensor)
Inlet Temperature
(External sensor)
Flue Gas Measurement
*2
Oxygen
*1
Carbon monoxide
*1
Carbon dioxide
Efficiency (Net or Gross)
*2
Efficiency High (C)
*2
Excess Air
*2
CO/CO
ratio
2
Pressure (differential)
Nominal range +80mbar
Maximum over range without
damage to sensor is
+400mbar
Pre-programmed Fuels
Storage Capacity
*1
Using dry gases at STP
*2
Calculated
Carbon Dioxide resolution is 0.01% below 1% measured value.
Range
o
0-600
o
0-50
o
0-600
0-21%
0-20ppm
21-2,000ppm nom
4,000ppm max for
15 mins
0-20%
*2
0-99.9%
0-119.9%
0-250%
0-0.999
+0.2 mbar
+1 mbar
+80 mbar
Natural gas, Propane, Butane, LPG, Light Oils (28/35 sec),
Wood Pellets
60 Combustion tests
20 Pressure & Temperature tests
20 Tightness tests
20 Temperature tests
20 Room CO tests
20 Commissioning tests
Resolution
o
C
0.1
C
o
C
0.1
C
o
C
0.1
C
0.1%
1ppm
0.1%
0.1%
0.1%
0.1%
0.0001
Maximum
0.001 mbar
<25mbar
Accuracy
o
+2.0
C
+0.3% reading
o
+1.0
C
+0.3% reading
o
+2.0
C
+0.3% reading
+0.3%
+3ppm
+5% reading
+0.3% volume
+1.0% reading
+1.0% reading
+0.2% reading
+5% reading
+0.005 mbar
+0.03 mbar
+3% of reading
Page 44