TIGHTNESS
PR1 20.01m
PR2 20.01m
TIME
When complete the display will show:
LOG
PR1 20.01m
PR2 19.98m
PR1 19.98m
PR2 19.97m
PRINT
Viewing / printing a logged Let-by and Tightness test
Select MENU / REPORT / TIGHTNESS / VIEW
Use
or
Press
/
Note:
The analyser's memory can store up to 20 tightness tests. Tightness tests
are logged automatically therefore the tightness section of the memory will
be full after the 20
can be performed the tightness section of the memory must be cleared. To
do this select MENU / REPORT / TIGHTNESS / DEL ALL / YES then press
.
119
01
to select the log number to be printed.
to print the test, (or send to PC via optional Wireless module).
th
tightness test is complete. Before the 21
Pressure at start of tightness test.
Real time pressure reading.
Tightness default time is 2 minutes. Can be
changed via "Menu".
The tightness test is automatically stored in
the memory.
Pressure at start of stabilisation test.
Pressure at end of stabilisation test.
Pressure at start of tightness test
Pressure at end of tightness test
Press
to print the complete test.
st
tightness test
Page 19