Cooling - IKA C 2000 basic Operating Instructions Manual

Calorimeter system
Table of Contents

Advertisement

3DJH 
 :DUUDQW\ FODLPV DQG OLDELOLW\
Please read these Operating Instructions carefully and in their entirety. IKA
consider itself responsible for the safety, reliability and performance of the device if
SDUWV FRQGXFWLQJ
The calorimeter system may only be opened by a customer service center.
HOHFWULFDO SRZHU
If service is required, we recommend that you contact our customer service depart-
ment.
Furthermore, we refer you to the applicable safety conditions and accident require-
ments.
IKA
of the device or impermissible modifications, repairs or renovations.
 6\VWHP SURSHUWLHV
The C 2000 calorimeter system is routinely used for determination of gross calorific
values of solid and liquid substances. The accessories of the system ensure cus-
tomized adaptation to laboratory tasks (see Chapter 13).
The system is characterized by the following features:
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com
the device has been operated in accordance with the Operating Instructions
only persons authorized by the manufacturer have performed repair and main-
tenance work on the device
only original parts and original accessories have been used in repair work.
®
is not responsible for damages or costs resulting from accident, improper use
Relieves you of routine tasks through automated measurement procedure
Integrated oxygen filling
Automatic detection of decomposition vessel (DV)
Operation without the cooling unit: Connection to a water faucet; temperature
range 12° C - 28° C; water consumption per measurement about 4 l; water pres-
sure 1 bar to max. 1.5 bar; at higher pressures, use C 25.
Operation with active cooling unit (for example IKA
Measurement and calculation of gross calorific value according to DIN 51900,
ISO 1928, ASTM D240, ASTM D4809, ASTM D5865, ASTM D1989, ASTM
D5468, ASTM E711
Calculation of net calorific value according to DIN 51900, ASTM D240,
ASTM D4809, ASTM D5865, ASTM D1989, ASTM D5468, ASTM E711
Measurement range: max. 40,000 J
(This corresponds to an increase in temperature in the inner vessel of about
5 K)
®
IKA
-WERKE C 2000 basic / control
 8VHU QRWHV
®
will only
®
KV 500, optional)
Ver. 04 08.07