&DXVH
Power failure
O
filling is not working
2
Insufficient quantity of oxygen in
the decomposition vessel.
Sample has low flammability
When the decomposition vessel
is suspended into the measuring
cell, the Bomb È display does
not change to START.
®
IKA
-WERKE C 2000 basic / control
5HPHG\
If no measurement was running, the
system can be restarted by turning it
off and back on again.
If the system was in the process of
performing a measurement, the pro-
cedure is as follows:
•
Turn the system off and back on
again and start in the normal,
proper manner.
•
If residual water is left in the in-
ner vessel, discharge it with
Maintenance menu (PSW\ ,9
•
Remove the decomposition ves-
sel, relieve the pressure and
prepare it again.
Examine:
•
the decomposition vessel to be
certain it is snugly in place (see
Section 7.4)
•
the condition of the seal on the
filling piston (see Section 11.5)
•
the seal of the decomposition
vessel C 5010/ 5012 Operating
Instructions)
Check the O
supply to the device
2
(30 bar).
Use combustion aids.
Check the following items:
•
The decomposition vessel was
not removed after the last meas-
urement.
•
No experiment has been pre-
pared.
•
The Maintenance menu is open.
•
There is a problem with the con-
tact spring (see Section 11.6).
•
The ignition wire is defective.
Ver. 04 08.07