Artisan Technology Group is your source for quality new and certified-used/pre-owned equipment SERVICE CENTER REPAIRS WE BUY USED EQUIPMENT • FAST SHIPPING AND DELIVERY Experienced engineers and technicians on staff Sell your excess, underutilized, and idle used equipment at our full-service, in-house repair center We also offer credit for buy-backs and trade-ins •...
Page 4
89 / 336 / CCE e 2006 / 95 CCE, in accordo ai seguenti regola- menti e documenti: DIN EN IEC 61 010-1 e DIN EN IEC 61 326-1. ® -WERKE C 2000 basic / control Ver. 04 08.07 Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com...
Page 5
Failure to observe this information may result in inaccurate measurement results. ® -WERKE C 2000 basic / control Ver. 04 08.07 Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com...
Turning on the system ..............6-9 Configuring the system ..............6-12 Turning off the system ..............6-19 ® -WERKE C 2000 basic / control Ver. 03 03.04 Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com...
Page 10
GDQJHU WKRVH ZRUNLQJ ZLWK LW RU WKLUG SDUWLHV You can obtain a copy of the pres- sure vessel regulation from Carl Heymann Verlag or Beuth Verlag. ® -WERKE C 2000 basic / control Ver. 04 08.07 Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com...
Page 11
® -WERKE C 2000 basic / control Ver. 04 08.07 Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com...
:DUUDQW\ In accordance with IKA warranty conditions, the warranty period is 12 months. For claims under the warranty please contact your local dealer. You may also send the machine direct to our works, enclosing the delivery invoice and giving reasons for the claim.
8VHU QRWHV 3DJH :DUUDQW\ FODLPV DQG OLDELOLW\ ® Please read these Operating Instructions carefully and in their entirety. IKA will only consider itself responsible for the safety, reliability and performance of the device if • the device has been operated in accordance with the Operating Instructions •...
Page 14
PC-operation of one or several calorimeters (CalWin software) • A sample rack can be connected • A printer can be connected ® -WERKE C 2000 basic / control Ver. 04 08.07 Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com...
Page 15
® -WERKE C 2000 basic / control Ver. 04 08.07 Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com...
Observe the instructions on han- dling oxygen given in Section 1 "For your safety”. ® -WERKE C 2000 basic / control Ver. 04 08.07 Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com...
, SW 17 Zuflussleitung Wasser Länge: 1,5 m Rückflussleitung Wasser Länge: 1,5 m Entleerschlauch Wasser Länge: 1m ® -WERKE C 2000 basic / control Ver. 04 08.07 Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com...
Page 18
PC, and keyboard) via the interfaces in the electronics unit is also shown on the display. ® -WERKE C 2000 basic / control Ver. 04 08.07 Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com...
Page 19
Pressure must be released from the decomposition vessel manually (venting button C 5010.6 or venting station C 5030). ® -WERKE C 2000 basic / control Ver. 04 08.07 Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com...
Page 20
For more information on operating the C 2000 control, please refer to the Operating ® Instructions for the C 5040 CalWin software package. ® -WERKE C 2000 basic / control Ver. 04 08.07 Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com...
1.5 bar. If necessary, you should ensure this by placing a pressure control valve (for exam- ple the IKA C 25) far enough forward in the circuit. DPRXQW RI Another important characteristic value for connecting a cooler/thermostat is the ILOOLQJ switch-off level (the minimum volume).
Page 22
® -WERKE C 2000 basic / control Ver. 04 08.07 Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com...
The solution obtained in this manner can now be examined with a suitable peripheral detection device for aqueous decomposition into the respective acid content. ® For more information on this subject, please contact IKA or your nearest authorized dealer. &RPSOHWH FRPEXVWLRQ To correctly determine the gross calorific value, it is of fundamental importance that the sample undergoes complete combustion.
FDSDFLW\ RI WKH V\VWHP PXVW EH GHWHUPLQHG IRU HDFK GHFRPSRVLWLRQ YHVVHO For more information on calibration, we refer you to the applicable standards. ® -WERKE C 2000 basic / control Ver. 04 08.07 Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com...
Page 26
(users in the USA). Please refer to chapter 6.2 for details how to connect to oxygen supplies. ® -WERKE C 2000 basic / control Ver. 04 08.07 Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com...
Page 27
7KH ZDWHU LQIORZ DQG RXWOHW OLQHV HDFK P ORQJ LQFOXGHG LQ WKH VFRSH RI VXSSO\ PXVW QHLWKHU EH H[WHQGHG QRU UHSODFHG ZLWK ORQJHU OLQHV ® -WERKE C 2000 basic / control Ver. 04 08.07 Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com...
Secure the water outlet line in place in the laboratory wash basin. Open the fitting valve completely. ® If you will be using the device in unsupervised operation, IKA recommends you use a standard commercial “water stop valve” in the water inflow line.
Page 29
Close the main valve on the oxygen supply when work is complete. Only carry out maintenance work when the system is depressurised. ® -WERKE C 2000 basic / control Ver. 04 08.07 Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com...
Page 30
The connection sockets are located on the rear wall of the device. When connecting the devices, please note the labeling on the connection wires. & VDPSOH UDFN VFDOH SULQWHU ® -WERKE C 2000 basic / control Ver. 04 08.07 Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com...
Page 31
7. /HIW 5LJKW 8S DQG 'RZQ DUURZ NH\ the arrow keys move the cursor within the input lines, menu windows, tables and protocols. ® -WERKE C 2000 basic / control Ver. 04 08.07 Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com...
Page 32
H[DPSOH RI D selection table GLDORJ ER[ input line H[DPSOH RI D ® -WERKE C 2000 basic / control Ver. 04 08.07 Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com...
Page 34
® -WERKE C 2000 basic / control Ver. 04 08.07 Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com...
Page 35
V\VWHP FKHFN VXFFHVVIXO After you confirm with 2., the device is ready for operation. ® -WERKE C 2000 basic / control Ver. 04 08.07 Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com...
Page 36
In this case a reference appears to the effect that there may be no valid calibrations available for the selected operating mode. ® -WERKE C 2000 basic / control Ver. 04 08.07 Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com...
Page 37
'RZQ DUURZ key or 2. you can open a menu window and after that a dialog box with 2. ® -WERKE C 2000 basic / control Ver. 04 08.07 Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com...
Page 38
If you confirm the dialog box with 2. the system clock and calendar will accept these values. ® -WERKE C 2000 basic / control Ver. 04 08.07 Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com...
Page 39
. Your entry is confirmed with “x“. Pressing the space bar again will delete the “x“ again. ® -WERKE C 2000 basic / control Ver. 04 08.07 Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com...
Page 40
This option is only valid for the C 2000, which is equipped with the C 60 calo- rimeter conversion kit. The standard decomposition vessels C 5010 and C 5012 must not be used in this mode. In the ([SORVLY mode, use exclusively the IKA ®...
Page 41
If you are working with different reference fuel, the gross calorific value of this fuel must be entered here. ® -WERKE C 2000 basic / control Ver. 04 08.07 Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com...
Page 42
( ) 0-NJ (YDOXDWLRQ configuration field • ( ) ',1,.$ ® The calculation modes used previously for IKA calorimeters are grouped under this evaluation procedure. ( ) $670 ' ' ' ' ' ( This procedure takes into account current US standards for combustion calorimetry of solid and liquid fuels, and wastes.
Page 43
® -WERKE C 2000 basic / control Ver. 04 08.07 Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com...
Page 44
The message also asks you to turn off the power switch for the calorimeter and cooler unit. ® -WERKE C 2000 basic / control Ver. 04 08.07 Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com...
Page 45
® -WERKE C 2000 basic / control Ver. 04 08.07 Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com...
Page 46
For systems with automatic DV detection, the calorimeter recognizes which vessel a measurement is being performed with and assigns its calibration parameters to it. ® -WERKE C 2000 basic / control Ver. 04 08.07 Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com...
Page 47
Readily combustible liquids with a low vapor pressure must not be come in direct contact with the cotton thread (for example tetramethyl dihydrogen disiloxan)! ® -WERKE C 2000 basic / control Ver. 04 08.07 Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com...
Page 49
Unscrew the cap screw and remove the cover by using the handle. RSHQLQJ WKH GHFRPSRVLWLRQ handle cover YHVVHO cap screw ® -WERKE C 2000 basic / control Ver. 04 08.07 Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com...
Page 50
If you use a scale in :LWK FRPEXVWLRQ DLG mode and previously transfer the weight of the combustion aid, the extraneous energy calculated from this information appears in this field. ® -WERKE C 2000 basic / control Ver. 04 08.07 Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com...
Page 51
Correction for the thermal energy from the cotton thread used as an ignition aid. A preset value of 50 J/0 J (without/with combustible crucible) appears here. If ® you use a different ignition aid instead of the IKA cotton thread, change this value as appropriate.
Page 52
This ensures that during the ignition process the burning thread will ignite the sample. ® -WERKE C 2000 basic / control Ver. 04 08.07 Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com...
Page 54
As many as 20 decomposition vessels per device can be used. As many measure- ments can be prepared as there are decomposition vessels available. ® -WERKE C 2000 basic / control Ver. 04 08.07 Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com...
Page 55
As soon as the system starts the experiment, the display shows the course of the temperature of the inner vessel over time as a graph. ® -WERKE C 2000 basic / control Ver. 04 08.07 Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com...
Page 56
The residue of combustion in the crucible, for example soot or ash, can also be wiped out with an absorbent and lint-free cloth. ® -WERKE C 2000 basic / control Ver. 04 08.07 Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com...
Page 57
® -WERKE C 2000 basic / control Ver. 04 08.07 Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com...
Page 58
Using the button &DOF in the window OLVW RI FDOLEUDWLRQV, carries out the acid correc- tion for the selected calibration. The calculation is dependent on the evaluation pro- cedure selected: ® -WERKE C 2000 basic / control Ver. 04 08.07 Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com...
Page 59
• Evaluation procedure DIN/IKA In the DIN/IKA evaluation procedure, the acid correction to ASTM D240 or ASTM D1989 can also be used for calibration. The results form is shortened ac- cordingly. The other calculation modes are not available for calibrations.
Page 60
Exit the dialog box with 2. The system calibration is thus complete and you can now continue with the determination of gross calorific values. ® -WERKE C 2000 basic / control Ver. 04 08.07 Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com...
Page 61
® -WERKE C 2000 basic / control Ver. 04 08.07 Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com...
Page 62
(isoperibolic 30 ° C), d25 (dynamic 25 ° C), d30 (dynamic 30 ° C) The experiment was stopped. ® -WERKE C 2000 basic / control Ver. 04 08.07 Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com...
Page 63
6LP The experiment was a simulation. (YDO The experiment was evaluated. – Code Eval: the DIN/IKA evaluation procedure was used. – Code ASTM: the ASTM D1989, D240, D5865, D4809, D5468, E711 evaluation procedure was used. ZDLW The fuel sample is in the crucible, entry of parameters is complete, and the experiment can be started.
Page 64
During and after the search pro- cedure, the header line shows the number of measurements found. ® -WERKE C 2000 basic / control Ver. 04 08.07 Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com...
Page 65
Open the (YDOXDWLRQ dialog box. You can reach this dialog box either through the ([SHULPHQWV menu in the header or through the (YDO. function key. H[SHULPHQW OLVW ® -WERKE C 2000 basic / control Ver. 04 08.07 Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com...
Page 66
The formulas and designations are those used in the ASTM standard. GLDORJ ER[ IRU $670 HYDOXDWLRQ ® -WERKE C 2000 basic / control Ver. 04 08.07 Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com...
Page 67
You can select from the following modes: ® -WERKE C 2000 basic / control Ver. 04 08.07 Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com...
Page 68
+\JU PRLVWXUH DQ The percentage of water from the hygroscopic moisture 9RODWLOHV UDZ The percentage of volatile components ® -WERKE C 2000 basic / control Ver. 04 08.07 Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com...
Page 69
The titrated quantity of 0.1 N barium hydroxide solution (titra- tion of distilled water with which the decomposition vessel was washed out after the experiment) ® -WERKE C 2000 basic / control Ver. 04 08.07 Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com...
Page 70
DIN 20 ml 0.05 N) added in the decomposition vessel 1D2+ The titrated quantity in ml (0.0866 N) ® -WERKE C 2000 basic / control Ver. 04 08.07 Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com...
Page 71
9RODWLOHV ZDI The percentage of volatile components in the “water and ash free” reference state ® -WERKE C 2000 basic / control Ver. 04 08.07 Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com...
Page 72
If you have confirmed the data with 2., a dialog box appears in which to enter the simulation parameters. ® -WERKE C 2000 basic / control Ver. 04 08.07 Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com...
Page 73
(YDOXDWLRQ dialog box (see Chapter 9 “Evaluation of gross calorific value determinations”) or you can convert the result into the desired reference state. ® -WERKE C 2000 basic / control Ver. 04 08.07 Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com...
Page 74
Water is pumped out of the inner vessel. Emptying is ended by time control (default: 90 sec). 7HPS ,QLW The temperature measurement is reinitialized. ® -WERKE C 2000 basic / control Ver. 04 08.07 Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com...
240 s is exceeded, the system will show an error message and the measurement will be broken off. (See also Section 12 „Troubleshooting“) ® -WERKE C 2000 basic / control Ver. 04 08.07 Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com...
To set the hose in place or to remove it, press the locking button on the water outlet connection. When emptying a cooler unit, please make note of the corresponding Operating Instructions. ® -WERKE C 2000 basic / control Ver. 04 08.07 Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com...
Page 77
The 7HPS HTXDOL] menu item results in overall temperature equalization for the calorimeter, which is particularly advantageous for the following messages. ® -WERKE C 2000 basic / control Ver. 04 08.07 Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com...
O ring 11 x 2 O ring 2 x 1.6 filling head centering ring cylinder screw GLVDVVHPEOLQJ WKH ILOOLQJ KHDG ® -WERKE C 2000 basic / control Ver. 04 08.07 Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com...
Page 79
If the power supply cord is replaced, one of comparable quality should be used. ® -WERKE C 2000 basic / control Ver. 04 08.07 Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com...
Page 80
(&RQI6HWWLQJV%RPE ,'). If the function cannot be restored, please contact the Technical Service Department. ® -WERKE C 2000 basic / control Ver. 04 08.07 Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com...
Page 81
(30 bar) • the combustibility of the sample; it may be necessary to add a combustion aid. ® -WERKE C 2000 basic / control Ver. 04 08.07 Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com...
Page 82
5.xx° C to 28° C before it is possible to continue working. ® -WERKE C 2000 basic / control Ver. 04 08.07 Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com...
Page 83
This number has nothing to do with operating time. When it is the storage capacity of the library. reached, the message above appears. ® -WERKE C 2000 basic / control Ver. 04 08.07 Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com...
Page 84
There is a problem with the con- tact spring (see Section 11.6). • The ignition wire is defective. ® -WERKE C 2000 basic / control Ver. 04 08.07 Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com...
Page 85
Maintenance menu: (PSW\ ,9. Water in the inner vessel cannot Contact the Technical Service be pumped out. Department. ® -WERKE C 2000 basic / control Ver. 04 08.07 Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com...
Page 86
Benzoic acid in tablet form (50 pieces) C 14 Disposable crucible (100 pieces) C 15 Paraffin strips (600 pieces) ® -WERKE C 2000 basic / control Ver. 04 08.07 Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com...
Page 87
® -WERKE C 2000 basic / control Ver. 04 08.07 Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com...
Page 88
1 x sample rack 1 x external monitor Connection for external monitor: Image frequency 115.74 Hz Line frequency 33.333 KHz ® -WERKE C 2000 basic / control Ver. 04 08.07 Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com...
Page 89
® -WERKE C 2000 basic / control Ver. 04 08.07 Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com...
Page 93
Development District Technology Park Fax: +49 7633 831-98 Tel.: 800 733-3037 510730 Guangzhou, China 47810 Petaling Jaya E-Mail: sales@ika.de Tel.: +1 910 452-7059 Tel.: +86 20 8222-6771 Selangor, Malaysia Fax: +1 910 452-7693 Fax: +86 20 8222-6776 Tel.: +60 3 7804-3322 E-Mail: usa@ika.net...
Page 94
Artisan Technology Group is your source for quality new and certified-used/pre-owned equipment SERVICE CENTER REPAIRS WE BUY USED EQUIPMENT • FAST SHIPPING AND DELIVERY Experienced engineers and technicians on staff Sell your excess, underutilized, and idle used equipment at our full-service, in-house repair center We also offer credit for buy-backs and trade-ins •...