3DJH
0HVVDJH
%RPE ³[´ DOUHDG\
DVVLJQHG
&RYHU QRW FORVHGRSHQ
:DWHU ILOOLQJ WLPH H[FHHGHG Fill state is not reached in the
1R WHPS LQFUHDVH
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com
&DXVH
The decomposition vessel is al-
ready assigned to a measure-
ment in progress. The measure-
ment procedure encountered
interference. The system dis-
plays the malfunction and inter-
rupts the measurement that is
starting.
The measurement cell cover is
not opening or closing com-
pletely.
The system displays the mal-
function and interrupts the
measurement.
inner vessel within 240 sec.
The system displays the mal-
function and interrupts the
measurement.
No increase in the temperature
after electrical ignition.
The system displays the mal-
function and interrupts the
measurement.
®
IKA
-WERKE C 2000 basic / control
7URXEOHVKRRWLQJ
5HPHG\
Turn the calorimeter off and back on.
Try to open the measuring cell cover
manually and then close it again. To
do this, in the Maintenance menu
execute the commands &ORVH 0&
and 2SHQ 0&.
If you cannot perform the function,
please get in touch with the Techni-
cal Service Department.
Repeat the experiment.
Check
•
the external water supply for cor-
rect functionality
•
the flowthrough on the device
To do this, select the &RROLQJ
ZDWHU item in the Maintenance
menu and measure the
flowthrough per unit of time on
the water outlet or check the
value F in the info-window (see
Section 11.3).
&RUUHFW YDOXHV DUH EHWZHHQ
PLQ OK DQG PD[ OK
If the setting is not correct,
please contact the Service De-
partment.
•
the dirt trap in the water inflow
line
Check for
•
correct position of the cotton
thread
•
a neat and tight contact of the
ignition wire
•
the O
supply (30 bar)
2
•
the combustibility of the sample;
it may be necessary to add a
combustion aid.
Ver. 04 08.07