IKA C 2000 basic Operating Instructions Manual page 83

Calorimeter system
Table of Contents

Advertisement

3DJH 
0HVVDJH
1R UHJXODWLRQ
(UURU ZDWHU VHQVRU
(UURU LJQLWLRQ FRQWDFW
6DPSOH UDFN LQYDOLG
1RW HQRXJK PHPRU\
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com
&DXVH
Temperature regulation cannot
reach a stable condition.
The temperature of the cooling
water is outside the permissible
range (12 C to 28 C).
The calorimeter is incorrectly
detecting a filled inner vessel.
The system displays the mal-
function and interrupts the
measurement.
This error message is displayed
if ignition capability can no longer
be ensured during a measure-
ment.
The experiment is stopped.
The actual assignment of the
sample rack does not agree with
the assignment that is adminis-
tered by the calorimeter.
The total number of all meas-
urements that the C 2000 can
administer during one run is lim-
ited to about 240.
This limit will be reached no
sooner than after 50 hours of
operating time. When it is
reached, the message above
appears.
®
IKA
-WERKE C 2000 basic / control
 7URXEOHVKRRWLQJ
5HPHG\
Check cooling water temperature
and correct if necessary.
Check the flow quantity and correct it
if necessary (see Section 11.3).
If there are any drops of water on the
water sensor, remove them (see
Section 11.6).
If the water level in the inner vessel
is higher than 3 cm, empty the inner
vessel.
If this does not resolve the problem,
please contact the Technical Service
Department.
Check:
the ignition wire
the ignition wire fastening
the electrodes in the decomposi-
tion vessel (the electrodes must
be tightened)
the contact spring on the filling
head (see Section 11.6).
See the Operating Instructions for
the C 5020 sample rack.
Turn the device completely off.
(in the System menu, menu item:
([LW)
1RWH
This number has nothing to do with
the storage capacity of the library.
Ver. 04 08.07