EasyLine EL3000 Series Commissioning Instructions

Status Messages

No. Status
Message
110 A S a
The instrument is booting.
116 A S a
The Profibus module is installed in the
wrong slot! The interface is therefore not
operative. Please install the Profibus
module in slot X20/X21.
119 A S iQ The configuration could not be loaded!
This instrument is therefore currently not
configured. Please load a configuration
using TCT.
120 F
a
The maintenance switch is ON.
121
aL The limit value has alarm status.
122 A S a
The IO module is defective.
123 A S a
Communication error while accessing the
IO module.
124
iQ The configuration data was corrupt! The
configuration was restored using the
backup data.
125
a
The limit value has alarm status.
126 W
a
The QAL3 data store is full. Please read
out the data.
127 W
a
The drift values exceed the QAL3 limits.
250 A S aQ The analyzer could not be found!
251 A S aQ The connection to the analyzer has been
lost!
252 A S aL The EEPROM data of the analyzer is
defective!
253 A S aL Communication with the analyzer is faulty! Check the connectors and cabling.
254 A S a
The boot program of the analyzer is
defective! Notify Service!
255 A S a
The program of the analyzer is defective!
Notify Service!
300 A S aL No new measured values from the
analog/digital converter.
301 A S a
The measured value exceeds the value
range of the analog/digital converter.
302 W
aQ The offset drift exceeds half the
permissible range.
303 A S aQ The offset drift exceeds the permissible
range.
Reaction/Troubleshooting
Install the Profibus module in slot X20/X21.
This instrument is therefore currently not
configured. Notify service if this occurs
again.
Exchange the IO module.
Exchange the IO module.
Export QAL3 data.
Repeat the calibration. Maintenance of
AMS required.
Check the connectors and cabling.
Check the connectors and cabling.
Check the configuration with TCT.
Notify Service.
Notify Service.
Notify Service.
Check the sample gas concentration.
Check connectors in the gas analyzer.
Notify Service.
Monitor the drift. The measured value is
valid as long as the drift remains below the
value specified in the measurement data.
As soon as the drift exceeds this value,
notify Service.
Notify Service. Carry out a basic
calibration (with TCT).