Shutting Down and Packing the Gas Analyzer

Shutting Down the Gas Analyzer

Shutting Down the Gas Analyzer
Fidas24: Gas Analyzer Shutdown
In the case of a temporary shutdown:
1
Shut off the sample gas.
2
Purge the gas lines and gas feed paths in the gas analyzer with dry
fresh air or nitrogen for at least 5 minutes.
3
Switch off the power supply to the gas analyzer.
In the case of a long-term shutdown, carry out the following in
addition:
4
Remove the gas lines from the gas analyzer ports. Tightly seal the
gas ports.
5
Disconnect the electrical leads from the gas analyzer.
In the case of a temporary shutdown:
1
Turn off the sample gas supply.
2
Purge sample gas line with nitrogen for at least 5 minutes from the
sampling point.
3
Set the gas analyzer to standby operation (see page 115). In case of
corrosive and flammable gases set the gas analyzer to standby
operation with detector purging.
4
Shut off combustion air supply and combustion gas supply.
In the case of a long-term shutdown, carry out the following in
addition:
5
Shut off instrument air supply.
6
Switch off the gas analyzer power supply.
7
Remove the gas lines from the gas analyzer ports. Tightly seal the
gas ports.
8
Disconnect the electrical leads from the gas analyzer.
Shutting Down and Packing the Gas Analyzer
131