No. Status
Message
359 A S a
The intensity measured at the detector
exceeds or falls below alarm value 2.
360 A S a
The filter wheel could not be initialized.
362 A S a
The calibration filter wheel cannot be
initialized.
378 A S aL The chopper wheel is blocked.
379 A S aL Chopper wheel speed not OK.
380 A S aL IR source element or electronics defective. Notify Service.
381 A S aL High voltage in the preamplifier defective.
382 A S aL Measured value is influenced by shocks.
390 A S aQ Failure of the power supply inside the
module. Notify Service.
397 A S a
The sensor of the temperature regulator is
defective.
398 A S aL No new measured values from the
analog/digital converter.
399 A S a
The measured value exceeds the value
range of the analog/digital converter.
400 A S a
A calculation error occurred during
calculation of the measured value.
401 W
a
The flow is above or below the upper or
lower alarm value 1.
402 A S a
The flow is above or below the upper or
lower alarm value 2.
403 A S a
The flow controller output variable is
beyond the valid range.
404 A S a
The temperature is above or below the
upper or lower alarm value 2.
411 F S a
The analyzer is in standby mode.
Reactivation in the Menu: Operation /
Pump/Standby / Fidas Standby.. Fidas
Restart.
412 F S a
Ignition failed. The analyzer must be
manually reactivated. Reactivation in the
Menu: Operation / Pump/Standby / Fidas
Standby.. Fidas Restart.
413 A S aL Auxiliary voltage failure in the analyzer
hardware.
414 F S a
The controlled variable of this controller is
below the permissible range (< 20%).
415 F S a
The controlled variable of this controller is
above the permissible range (< 90%)
500
iQ An internal calibration error has occurred.
501
iQ The requested functional capability is not
available in the instrument.
Status Messages, Troubleshooting
Reaction/Troubleshooting
Notify Service.
If the status message is active for longer
than 5 minutes, the filter wheel is blocked.
If the status message is active for longer
than 5 minutes, the calibration filter wheel
is blocked.
Visual check. Notify Service.
Notify Service.
Notify Service.
Notify Service.
Notify Service
Notify Service.
Check the sample gas concentration.
Check connectors in the gas analyzer.
Notify Service.
Switch the power supply off and on. Notify
Service.
Check sample gas path. Notify Service.
Check sample gas path. Notify Service.
Notify Service.
Notify Service.
Restart of the Fidas24.
Check process gases.
Restart of the Fidas24.
Notify Service.
Check the process gases and connection
lines. If required, restart of the Fidas24.
Notify Service.
Repeat the calibration. Check the
configuration with TCT. Notify Service.
Check the configuration with TCT.