ABB EasyLine Series Instructions For Installation Start-Up And Operation page 126

Continuous gas analyzers
Table of Contents

Advertisement

126
EasyLine EL3000 Series Commissioning Instructions
No. Status
Message
321 A S a
The temperature of the detector is below
the minimum temperature.
322 A S a
The flame is out.
323 A S a
The analyzer is in the fail-safe status.
324 W
a
The temperature is above or below the
upper or lower alarm value 1.
325 W
a
The temperature is above or below the
upper or lower alarm value 2.
326 A S aL No new measured values from the
analog/digital converter.
327 A S a
The measured value exceeds the value
range of the analog/digital converter.
328 A S a
A calculation error occurred during
calculation of the measured value.
329 W
a
The pressure is above or below upper
and/or lower limit value 1.
330 W
a
The pressure is above or below upper
and/or lower limit value 2.
331 A S a
The pressure regulator output variable is
beyond the valid range.
342 W
a
The flow falls below limit value 1
343 A S a
The flow falls below limit value 2.
357 A S a
Limas motor optimization is in progress.
358 W
a
The intensity measured at the detector
exceeds or falls below alarm value 1.
Reaction/Troubleshooting
Status message during the warm-up
phase. If the status message appears
following the warm-up phase: Check the
overtemperature cutout and replace if
necessary.
Status message during the warm-up
phase. If the status message appears
following the warm-up phase: Check gas
supply, heater plug.
Causes: Flame temperature > detector
setpoint + 220 °C, hardware fault, Pt-100
line break or short circuit.
Switch power supply off and on again
after 3 seconds. If this status message
appears again, notify Service.
Note: Fail-safe-status means: Heater off,
combustion gas valve closed, instrument
air valve closed, housing purge on, zero
gas valve open.
Status messages during the warm-up
phase. If the status message appears
following the warm-up phase: Check
whether the permissible ambient
temperature range is being maintained.
Check the connecting leads and
connectors. Check the fit of the leads in
the wire end ferrules. Check the
overtemperature cutout and replace if
necessary.
Notify Service.
Notify Service.
Notify Service.
Fidas24: Check the supply gas pressures.
Fidas24: Check the supply gas pressures.
Fidas24: Check the supply gas pressures.
Check sample preparation.
Alarm value 1 = 25 % of MRS.
Check sample preparation.
Alarm value 2 = 10 % of MRS.
Automatic calibration is interrupted and
disabled.
If the status message is active for longer
than 5 minutes, the filter wheel is blocked.
Notify Service.