EasyLine EL3000 Series Commissioning Instructions
Ignition of the flame
12 The flame ignites automatically:
Temperature
of detector and sample gas
connection > 150 °C?
yes
P
ressure values
operating gases adjustable
to set-point?
yes
Glow plug
switched on
Combustion air pressure
set to start value
Combustion air pressure
raised one step
Flame on? (Flame temp.
> detector temp. + 30 °C)
yes
Glow plug
switched off
Pressure values of operating
gases adjusted to set-point
Fidas24 is ready to measure
after warm-up phase
Depending on the number of ignition attempts, the flame can take up
to 10 minutes to ignite.
Depending on the length of the combustible gas supply line, it is
possible that not enough combustible gas is available to ignite the
flame at first during the initial start-up of the gas analyzer. In this
case, the ignition of the flame must be restarted in the Fidas
Restart menu (see page 115). The temperature of the flame is also
displayed in this menu option; the flame is regarded as "on" when the
flame temperature is at least 30 °C higher than the detector
temperature.
The actual start-up of the gas analyzer is concluded with the ignition
of the flame.
no
Operating temperature
of
no
Check pressure values
Fidas24 goes to standby mode
and is ready for restart
no
ramp height reached?
Number of ignition attempts
Combustion gas pressure
Fidas24 goes to standby mode
and is ready for restart
Status message:
Status message:
Pressure error
of operating gases
Combustion air
yes
raised one step
raised one step
Number of ignition
attempts > 10?
yes
Glow plug
switched off
no
no