Magnos206
Magnos27
Caldos27
Sample Gas Inlet Conditions
Temperature
+5...+50 °C
Positive pressure
2...100 hPa
Flow
30...90 l/h
Flammable Gases
The analyzer is suitable for the measurement of flammable gases in a
non-hazardous environment. The special requirements for the
measurement of flammable gases (see page 38) must be complied with.
Corrosive Gases
If the sample gas contains Cl
the analyzer may only be used if the sample gas composition has been
taken into account by the manufacturer for the configuration of the
analyzer.
Sample Gas Inlet Conditions
Temperature
+5...+50 °C
Positive pressure
2...100 hPa
Flow
20...90 l/h
Flammable Gases
The analyzer may not be used for the measurement of flammable gases.
Sample Gas Inlet Conditions
Temperature
+5...+50 °C
Positive pressure
2...100 hPa
Flow
10...90 l/h, min. 1 l/h
Pressure drop
< 2 hPa at 60 l/h N
Flammable Gases
The analyzer is suitable for the measurement of flammable gases in a
non-hazardous environment. The special requirements for the
measurement of flammable gases (see page 38) must be complied with.
Corrosive Gases
If the sample gas contains Cl
components, the analyzer may only be used if the sample gas
composition has been taken into account by the manufacturer for the
configuration of the analyzer.
Preparation for Installation
, HCl, HF or other corrosive components,
2
2
, HCl, HF, SO
, NH
, H
2
2
3
21
S or other corrosive
2