Manual Calibration: Execution

Menu Path
Wait until the Warm-up Phase has Ended
Operation
Calibration
The gas analyzer may not be calibrated until the warm-up phase has
been concluded.
Duration of the warm-up phase:
Uras26: approx. ½ hour without thermostat, approx. 2 hours with
thermostat
Limas23: approx. 2 hours
Magnos206: approx. 1 hour
Magnos27: approx. 2–4 hours
Caldos27: approx. ½ hour
Fidas24: approx. 2 hours
NOTES
A zero point calibration must always be carried out before an end-point
calibration. A zero-point calibration can also be carried out on its own.
Manual calibration cannot be carried out when an automatic calibration is
running.
Gas Analyzer Calibration
Manual Calibration
113