Gas Connections for Limas23 (Model EL3040)
Limas23: Gas Connections
13 Sample gas inlet
14 Sample gas outlet
Purge gas inlet for case
11
12 Purge gas outlet for case
Design:
Note:
1/8 NPT female thread (stainless steel 1.4305/SAE 303)
Connection of flexible tubes: Straight screwed fittings (PP) with nozzles for flexible
tubes with inside diameter = 4 mm (supplied)
Connection of pipelines: Threaded connections (not supplied)
The pressure sensor (standard) and the oxygen sensor (option) are connected
internally in the outlet of the sample cell.
Installing the Gas Analyzer
55