EasyLine EL3000 Series Commissioning Instructions
Calibrate the Sample Component Manually
Zero-point Calibration
1
Select menu
Zero point/Single point
2
Select individual sample component or "All" (according to the
configuration in the ECT dialog "Manual Calibration").
3
Check zero point set point
4
Connect zero reference gas (if it is not automatically connected).
5
Start calibration as soon as the measured value display is stable.
6
Store calibration or repeat calibration
End-point Calibration
1
Select menu
Span
2
Select sample component.
3
Check span set point
4
Connect span gas (if it is not automatically connected).
5
Start calibration as soon as the measured value display is stable.
6
Store calibration or repeat calibration.
End-point calibration with calibration cell (option with Uras26 and
Limas23)
1
Select menu
Calibration Cell
2
Select sample component or "All" (only possible with Uras26).
3
Connect zero reference gas (if not already connected).
4
Start calibration as soon as the measured value display is stable.
5
Store calibration or repeat calibration.
4
The parameterized test gas concentration is displayed. If the setpoint is altered here, the
parameterized test gas concentration is overwritten.
5
A calibration may have to be repeated if the measured value is not stable after initiation of
the calibration. The repeated calibration is based on the measured value obtained in the
preceding calibration.
6
The parameterized test gas concentration is displayed. If the setpoint is altered here, the
parameterized test gas concentration is overwritten.
4
and adjust if required.
5
.
.
6
and adjust if required.
.
.