Proceed as follows to activate deep sleep mode:
1 Select "Device Settings" in the main menu and confirm the selection
2 A list of the parameters available for the current MyMode is displayed The entry "Deep sleep" is at the end of
this list; select it and press the enter button
3 A confirmation prompt is displayed Press the enter button again
• Feedback from the prosthesis: 1 short beep signal
Proceed as follows to wake the knee joint using the remote control:
1 Activate the remote control by pressing the enter button and 
2 Select the "Device Power On" menu entry and press the enter button to confirm
3 The knee joint is reactivated after 20 seconds at the latest with a single beep and vibration signal The knee
joint is then in basic mode again
Alternatively, the knee joint can be woken by attaching the charger
INFORMATION
Deep sleep mode is not available if the joint is in safety mode
6.8 Querying the Prosthesis Status
INFORMATION
Advise your patients of the information in this section
1 Selecting and confirming "Status" displays the following:
• The total pedometer value ("Step")
• The daily pedometer value ("Trip")
• The charging status of the prosthesis in percent ("Batt")
• The estimated remaining operating time of the prosthesis in hours during active use and at rest (eg sitting
function) („Stb/Act" for Standby/Active)
• The next service due date
2 Select "Trip" and press enter to reset the daily pedometer