Remote control
6.2 Control Elements of the remote control
INFORMATION
Advise your patients of the information in this section
3
6.3 Initial Connection between the remote control and the knee joint (Pairing)
1 Press the enter button to activate the remote control The start-up screen with the date, time and charging
state of the remote control is displayed Without additional input, the remote control turns off automatically
after 5 seconds
2 Press to attempt to connect the device The remote control searches for active prostheses
3 When a knee joint is found (serial number is shown), press enter to confirm If several knee joints are found,
the serial number of the desired knee joint has to be selected using the navigation keys (display: serial num-
ber) The configuration data from the knee joint is then loaded
INFORMATION
This process can take a few minutes
4 The main menu is displayed
After successful initial connection with the knee joint, the remote control always connects automatically after acti-
vation (press the enter button first, then ) No further steps are required
6.4 Status Bar
The status bar shows:
1 Charging state of the remote control
2 Charging state of the knee joint with percentage indicator
3 Current MyMode
4 In case of a deviation from normal knee joint functionality
(eg when the charger is connected), an exclamation mark
is shown in the status bar at the top right The current My-
Mode is not displayed in some of these situations
20 | Ottobock
1 LCD screen
2 Enter button
1
3 Navigation buttons: left, right, up, down
4 Charging receptacle
2
3
4
Genium X3 - Bionic Prosthetic System