INFORMATION
Hold the joint still while it is performing the self-test, ie immediately after removing the charger Otherwise,
error messages may occur They can be rectified by connecting and removing the charger
INFORMATION
Charging is no longer possible when the distance between the transmitter and receiver is greater than 2 mm
However, charging should not be a problem if there is a sock or Superskin between transmitter and receiver of
the charger (See also the charging process table)
Charging process:
LEDs
0
Charging
0%-10%
status
Charging process table:
Status
AC Adapter is connected, but the
charger is not connected to the
knee joint
Charger connected to knee joint
Charger connected to knee joint,
bad connection
a) Charger overheating
b) knee joint overheating/too cold
c) Connection may be improved by
reducing the distance between the
charger and receiver
Charger fault (error may be rectified
by disconnecting and reconnecting
the AC Adapter)
* This display turns off automatically after approximately one minute to avoid bothersome light at night The char-
ging process is not affected
The charging status of the knee joint is also shown on the display of the remote control
In addition, the remaining operating time is indicated to the patient by knee joint vibration signals once the remai-
ning battery capacity reaches approximately 25% (see table below)
vibration signal of the knee joint
3 x vibration
5 x vibration
10 x vibration and beeping
INFORMATION
When the knee joint is fully charged (all 5 LEDs at the side light up (Fig 3)), the maximum period of use for the
knee joint before the next charging process is up to 5 days, depending on patient activity However, we recom-
mend charging the knee joint every night
Genium X3 - Bionic Prosthetic System
1
2
10%-30%
30%-50%
Status LED
Continuous green
Purple pulsing (4-second-
sond cycle)
*
Weak purple
*
Continuous yellow The
joint is not charged
Flashing red The joint is
not charged
Charging status
25 %
10 %
5 %
Alignment and Adjustment of the Prosthesis
3
4
50%-70%
70%-90%
Charging status indicator
All five LEDs off
Charging status indicator (green)
according to the number of LEDs lit
Charging status indicator (green)
according to the number of LEDs lit
a) 2nd and 4th LED illuminated
a) 1st, 3rd and 5th LED illuminated
c) 3rd LED illuminated
(See Section 21, Fig 3)
LEDs indicate an error code
Remaining operating time
approx 24 h
approx
6 h
0 h
5
>90%
Ottobock | 13