Otto Bock Genium X3 Instructions For Use Manual page 28

Bionic prosthetic system
Hide thumbs Also See for Genium X3:
Table of Contents

Advertisement

Important Information for Patients
Type of error
Minor error
Medium error
AXON Tube Adapter
disconnected
Major error
Total failure
Charging Status 0%
Error code on the
display of the Remo-
te Control
In addition to the basic mode and the MyModes, the knee joint has a range of safety modes which provide various
levels of functionality depending on the type and severity of the errors As soon as the knee joint detects a fault,
at least five long beep signals and 5 slow pulsating vibration signals are emitted The knee joint activates a safety
mode and remains in this status until the error is rectified (provided the battery capacity is sufficient)
The OPG function, Stairs and Obstacles function, Intuitive Stance function and Sitting function are always deac-
tivated in the safety mode; there are also various levels of basic functionality:
• Continuous high flexion resistance
• Continuous stance phase flexion resistance with the ability to initiate the swing phase
Depending on the type of error, swing phase initiation locking may or may not be active while walking backwards
You must therefore exercise caution when walking backwards after an error
The swing phase control and stance phase extension resistance may be available or not depending on the type
of error
If safety mode was triggered due to a empty battery (see Section 72 "Charging Status"), charging the prosthesis
resets it to basic mode In this case, the resistance value configured in the X-Soft adjustment software is activated
There is also an overheating mode to protect the knee joint against excessive temperatures In this safety mode,
stance phase resistance increases as the temperature rises – especially in order to slow down walking downhill
so that overheating is not allowed to worsen Swing phase control is also progressively reduced
Overheating mode is indicated by brief vibration every 5 seconds
Basic mode is reactivated once the temperature has fallen sufficiently
28 | Ottobock
Event
eg service interval has
expired
eg a sensor (tube adap-
ter, triple sensor, ) is not
operating
Selftest failure (if signals
occur after removing the
charger)
AXON Tube Adapter has
been disconnected while
the knee joint is activated
and the inductive charger
is not attached
AXON Tube Adapter not
connected while removing
inductive charger
eg failure of the valve
drive
Selftest failure (if signals
occur after removing the
charger)
No electronic functionality
User restriction
No restriction
Walking possible with
restrictions (activation of
safety mode)
• High stance resistance
• Swing phase no longer
possible
• Activation
of
safety
mode (dependant on
the remaining functio-
nality)
• Swing phase no longer
possible
Genium X3 - Bionic Prosthetic System
Required action
Consult Ottobock
Restart the knee joint by
attaching and removing
inductive charger (up to
three times);
If error remains, consult
Ottobock immediately
Connect the AXON Tube
Adapter
Restart the knee joint by
attaching and removing
inductive charger (up to
three times);
If error remains, consult
Ottobock immediately
Consult Ottobock imme-
diately
Charging required
Consult Ottobock immedi-
ately (Error Code number
required)