Mymode Motion Switching; 51 Switching To Mymodes Without A Remote Control 18 - Otto Bock Genium X3 Instructions For Use Manual

Bionic prosthetic system
Hide thumbs Also See for Genium X3:
Table of Contents

Advertisement

MyMode Motion Switching

5 MyMode Motion Switching

INFORMATION
Advise your patients of the information in this section
5.1 Switching to MyModes without a remote control
If switching by bouncing up and down has been activated in X-Soft, you can switch without a remote control by
bouncing up and down on the forefoot three times in a second, with the foot remaining in contact with the ground
at all times and the prosthetic leg extended While doing so, the forefoot has to bear approximately 40% of the
total body weight at peak load and less than approximately 25% of the body weight at the lowest load However,
the load may not be taken off the foot fully
3 ×
in 1 second
After the audible signal sounds (one long beep and a long vibration signal), your weight must be taken off the
prosthesis entirely for at least one second with the knee extended (no ground contact) Then the knee joint swit-
ches to the first MyMode and emits two short Beep and vibration signals
The procedure for switching to the second MyMode is the same, except for bouncing up and down four times In
this case, the switch is indicated by three short Beep and vibration signals
Bouncing at least three times to return to basic mode from any MyMode The switch is confirmed by a short Beep
and vibration signal
If only triple bouncing is activated but quadruple bouncing is not, you can access the first MyMode via any number
of bouncing motions greater or equal to 3
INFORMATION
Switching back to basic mode by bouncing up and down three or more times cannot be deactivated Thus, you
can access basic mode even if switching via bouncing is not enabled in X-Soft
INFORMATION
Returning to basic mode is also possible by connecting the battery charger
CAUTION
Feedback is disabled in mute mode. When the mute mode is active, no feedback will be given when swit-
ching into a MyMode or back to basic mode If switching was unsuccessful, the behaviour of the knee joint may
be unexpected
This may result in a risk of falling Make sure that the switching process has been successful before continuing
to use the knee joint
18 | Ottobock
1 second
1
2
Genium X3 - Bionic Prosthetic System
3

Advertisement

Table of Contents
loading